The Problem

The v1 CLARIOstar reads absorbance and fluorescence, but not luminescence, although the reader reports a luminescence measurement channel. Luminescence assays (luciferase reporters, ATP viability, BRET) cannot run on it through PyLabRobot.

PR Content

  • Luminescence.measure: a point read of each well from the top, with enhanced dynamic range, through one emission band of the emission monochromator.
  • Its settings: emission_wavelength / emission_bandwidth (580 / 80 nm), gain (3500), focal_height (None: a quarter up the well, as for fluorescence), normalization_time (1.00 s), measurement_interval (0.80 s), plus MeasurementSettings (settling, corner, reading direction) and ShakeSettings as the other modes take them.
  • The emission limit follows the photomultipliers, as for fluorescence but the other way round: the extended range (up to 840 nm) needs the extended-red photomultiplier on the luminescence path - fitted, and crossed over on a dual-photomultiplier reader; otherwise 740 nm.
  • pylabrobot.lib.luminescence.LuminescenceResult, the same for every plate reader; CLARIOstarLuminescenceResult adds the gain.
  • driver.luminescence / CLARIOstarDevice.luminescence, built at setup only when the reader reports a luminescence channel.
  • FluorescenceConfiguration.maximum_chromatics is renamed maximum_fluorescence_channels.
  • Comments, docstrings and test names across the CLARIOstar driver are reworded; nothing changes on the wire. The test data files are renamed *_ground_truth.json.gz.
  • Until a luminescent sample has been read on hardware, measure warns that luminescence is untested.

🤖 Generated with Claude Code

@BioCam
BioCam force-pushed the bmg-clariostar-v1-luminescence branch 7 times, most recently from 6a5e771 to 4419b2e Compare October 22, 2026 06:25
@BioCam
BioCam marked this pull request as ready for review October 22, 2026 08:28
@BioCam
BioCam requested a review from a team as a code owner October 22, 2026 08:28
@BioCam
BioCam force-pushed the bmg-clariostar-v1-luminescence branch from 4419b2e to 31dc2c2 Compare October 22, 2026 08:30
- `Luminescence.measure` reads the plate's luminescence from the top, each well at its centre,
  with enhanced dynamic range, through one emission band of the monochromator.
- The emission band, gain, focal height, normalization time, measurement interval, settling,
  reading direction, corner and shaking are sent.
- The emission limit follows the photomultipliers: up to 840 nm with the extended-red one on the
  luminescence path, 740 nm without.
- `pylabrobot.lib.luminescence.LuminescenceResult`; `CLARIOstarLuminescenceResult` adds the gain.
- `driver.luminescence` and `CLARIOstarDevice.luminescence` exist when the reader reports a
  luminescence channel.
- `FluorescenceConfiguration.maximum_chromatics` is renamed `maximum_fluorescence_channels`.
- Comments, docstrings and test names reworded, nothing changes on the wire; the test data files
  are renamed `*_ground_truth.json.gz`.

Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
@BioCam
BioCam force-pushed the bmg-clariostar-v1-luminescence branch from 31dc2c2 to 262b2d5 Compare October 22, 2026 09:37
@BioCam
BioCam merged commit 30b87b9 into PyLabRobot:main Oct 22, 2026
27 checks passed